Fluke HART SCIENTIFIC 1502A Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Thermomètres Fluke HART SCIENTIFIC 1502A. Fluke HART SCIENTIFIC 1502A User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - User’ s Guide

1502AThermometer ReadoutUser’ s GuideRev. 722101 ENGHart Scientific

Page 2 - Rev. 722101

Keep the probe wires clean and away from fluids.1.3 Authorized Service CentersPlease contact one of the following authorized Service Centers to coordi

Page 3 - Table of Contents

Fluke South East Asia Pte Ltd.Fluke ASEAN Regional OfficeService Center60 Alexandra Terrace #03-16The Comtech (Lobby D)118502SINGAPOREPhone: +65 6799-

Page 4

2 IntroductionThe 1502A is a low-cost high-accuracy digital thermometer readout designed tobe used with 25Ω and 100Ω RTDs and SPRTs. Its unique combin

Page 5

3 Specifications and EnvironmentalConditions3.1 SpecificationsResistance Range0Ωto 400Ω, auto-rangingResistance Accuracy, one year10Ωto 20Ω: 0.0005Ω20

Page 6

Weight2.2 lb. (1.0 kg)SafetyOVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 perIEC 1010-11Accuracy specifications apply within the recommen

Page 7

4 Quick StartThis section briefly explains the basics of setting up and operating your 1502Athermometer readout.4.1 UnpackingUnpack the thermometer ca

Page 8 - 1.2 Safety Information

to pins 4 and 5. (Pins 1 and 5 source current and pins 2 and 4 sense the poten-tial.) If a shield wire is present it should be connected to pin 3.A tw

Page 9 - Cautions

5 Parts and Controls5.1 Front Panel ButtonsSee Figure 2.The front panel buttons are used to select units of measurement, access operat-ing parameters,

Page 10 - User’s Guide

5.2 Rear PanelSee Figure 3.Serial Port - The DB-9 connector is for interfacing the thermometer to a com-puter or terminal with serial RS-232 communica

Page 11 - Authorized Service Centers

6 General OperationThis section explains basic operation of the 1502A Thermometer.6.1 Selecting UnitsTemperature can be displayed in degrees Celsius (

Page 12 - 2 Introduction

Limited Warranty & Limitation of LiabilityEach product from Fluke Corporation, Hart Scientific Division ("Hart") is warranted to be free

Page 13 - Conditions

1502A Thermometer ReadoutUser’s Guide16MenuSample Comm (Cal)ProbeSet clockSet filterSet probe type Enter passwordSet time stampSet currentSet coeffici

Page 14 - 3.2 Environmental Conditions

when all the digits are correct. If the password is entered correctly the first pa-rameter in the menu will appear.6.4 Selecting the Probe Characteriz

Page 15 - 4 Quick Start

digit will flash. Use the U and D buttons to change a digit. Once the sign anddigits are correct, press Enter to accept the number. If you decide to c

Page 16 - Probe Connector

Example 1:A PRT was calibrated to ITS-90 and its calibration certificate states values forcoefficients Rtpw, a4, b4, a8, and b8. Set the 1502A paramet

Page 17 - 84.981 C

Example 3:A PRT was calibrated to ITS-90 and its calibration certificate states values forcoefficients R(273.16K), a6, b6, c6, and d. Set the 1502A pa

Page 18 - 5.2 Rear Panel

6.4.5 IPTS-68 ConversionThe IPTS-68 characterization converts resistance to temperature according tothe International Practical Temperature Scale of 1

Page 19 - 6 General Operation

Compare this temperature with the temperatures listed on the probe calibrationreport to verify that the coefficients you entered are correct.6.5 Filte

Page 20 - Sample Comm (Cal)Probe

7 Digital Communications InterfaceRemote communications allows an external device, such as a computer, to com-municate with the 1502A to obtain measur

Page 21 - +”). Use the

7.1.2 Automatic Transmission of MeasurementsThe 1502A can be programmed to automatically send measurements to a re-mote printer or terminal. The trans

Page 22 - E -04

to the time stamp parameter. The display will briefly show “ti Sta” then thetime stamp state which is either ON or OFF. Use the U and D buttons tochan

Page 23

Table of Contents1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Symbols Used . . . . . . .

Page 24

7.2.1 Setting the AddressThe IEEE-488 bus requires that each device has a unique address. The defaultaddress of the 1502A is 22 but can be changed if

Page 25 - E -04

277 Digital Communications InterfaceRemote CommandsCommand DescriptionMeasurement CommandsT read measurement (includes label, unit, and time)F[ETCH?]

Page 26 - 6.7 Power Saver

7.3.1.1 Reading TemperatureThe most recent temperature measurement can be read using the followingcommand:T<EOS> reads the most recent measureme

Page 27 - 7.1 Serial Interface

7.3.1.2 Automatically Transmitting MeasurementsBy setting the sample period, the 1502A can be programmed to automaticallytransmit measurements from th

Page 28

7.3.2.1 Selecting the CharacterizationThe following commands can be used to select the probe characterization andcoefficients:P=90<EOS> selects

Page 29 - 7.2 GPIB Interface

7.3.3.1 Setting the FilterThe filter helps to reduce variations in the measurements. The filter can be setremotely using the command:FI=<value>&

Page 30 - 7.3 Remote Commands

LF=OF<EOS> disables linefeed7.3.5 Calibration CommandsThe following commands are used in calibrating the instrument.7.3.5.1 Entering the Passwor

Page 31

ver.mmmmm,v.vvThe m’s represent digits of the model number. The v’s represent the digits ofthe firmware version number. As an example, if the version

Page 32 - Command List Continued

8 Calibration ProcedureThe 1502A uses a three-point calibration scheme with a quadratic polynomialcorrection function to maintain the accuracy of its

Page 33 - Remote Commands

power. The *C0=<value><EOS>, *C1=<value> <EOS>, and*C4=<value><EOS> commands can be used to set the values of the

Page 34 - 7.3.3 Sample Commands

7 Digital Communications Interface . . . . . . . . . . . . . . . 237.1 Serial Interface . . . . . . . . . . . . . . . .

Page 35 - 7.3.4 Communication Commands

9 Maintenance•The calibration instrument has been designed with the utmost care. Easeof operation and simplicity of maintenance have been a central th

Page 36 - 7.3.6 Other Commands

10 TroubleshootingIn case you run into difficulty while operating the 1502A, this section providessome suggestions that may help you solve the problem

Page 37

•Stem conduction error. The problem may be that the actual temperatureof the sensor is not what you expect. This is often the result of stem con-ducti

Page 38 - 8 Calibration Procedure

10.1.1.1 Immunity TestingThe instrument was tested to the requirements for industrial locations. This al-lows the instrument to be used in all types o

Page 39 - 8.2 Calibration Procedure

iiiFiguresFigure 1 Connecting a four-wire probe . . . . . . . . . . . . . . . . . . . . . . 12Figure 2 1502A Front Panel . . .

Page 40 - 9 Maintenance

ivTablesTable1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1Table 2 Matching Certificate Values to

Page 41 - 10 Troubleshooting

1 Before You Start1.1 Symbols UsedTable 1 lists the symbols that may be used on the instrument or in this manualand the meaning of each symbol.Symbol

Page 42 - 10.1 CE Comments

Symbol DescriptionOnCanadian Standards AssociationOVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re-fers to the level of Imp

Page 43 - CE Comments

Before initial use, or after transport, or after storage in humid or semi-humidenvironments, or anytime the instrument has not been energized for more