Fluke 434 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Acoustique Fluke 434. Fluke 434 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 122
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
®
Fluke 433/434
Three Phase Power Quality Analyzer
Users Manual
EN
August 2004
© 2004 Fluke Corporation, All rights reserved. Printed in the Netherlands
All product names are trademarks of their respective companies.
Vue de la page 0
1 2 3 4 5 6 ... 121 122

Résumé du contenu

Page 1 - Fluke 433/434

® Fluke 433/434 Three Phase Power Quality Analyzer Users Manual EN August 2004 © 2004 Fluke Corporation, All rights reserved. Printed in the Net

Page 2

Fluke 433/434 Users Manual 1-4 Shipment Note The following items are included in your Analyzer Kit: Note: When new, the Analyzer’s rechargeable NiMH

Page 3 - Table of Contents

Fluke 433/434 Users Manual 20-2 Proceed as follows to activate the Advanced Functions: • Press the SETUP key to enter the SETUP entry menu. • Press

Page 4 - ii

Tips and Maintenance Troubleshooting20 20-3 Optional Accessories. Item Ordering Code Advanced Functions for Fluke 433 (Interharmonics, Transien

Page 5 - iii

Fluke 433/434 Users Manual 20-4 Screen remains black. Make shure that the Analyzer is on: at power-on you should hear a double beep. If the screen rem

Page 6 - iv

21-1 Chapter 21 Specifications Introduction Performance Characteristics Fluke guarantees the properties expressed in numerical values within the t

Page 7

Fluke 433/434 Users Manual 21-2 Electrical Measurements The following specifications of the instrument are verified using the “implementation verifica

Page 8

Specifications Electrical Measurements21 21-3 RMS Voltage Selected Nominal Voltage (Vnom) Measurement Range (CF ≤ 1.4 at full scale) Resolution

Page 9 - Declaration of Conformity

Fluke 433/434 Users Manual 21-4 Measurement Measurement Range Resolution Accuracy Relative Amplitude 0.0 ... 100.0 % 0.1 % ± 0.1 % ± n x 0.1 %,

Page 10 - Shipment Note

Specifications Electrical Measurements21 21-5 Voltage Interrupts Settings Adjustment Range Resolution Additional Information Interrupt Threshol

Page 11 - Contacting a Service Center

Fluke 433/434 Users Manual 21-6 Measurements Measurement Range Resolution Accuracy RMS voltage in Transient Function 10 ... 1000 Vrms 1 Vrms ± 2.

Page 12

Specifications Electrical Measurements21 21-7 RMS Current Selected Clamp Sensitivity Measurement Range ( CF ≤ 2.8 at full scale) Resolution Acc

Page 13 - General Aspects

General Aspects Contacting a Service Center 1 1-5 # Description 1 Power Quality Analyzer 2 Decal Set for Input Sockets 3 Hang Strap 4 Alliga

Page 14

Fluke 433/434 Users Manual 21-8 Inrush Current Settings Adjustment Range Resolution Additional Information Inrush Threshold level 0 ... 999 % of Ino

Page 15 - About This Manual

Specifications Electrical Measurements21 21-9 POWER MEASUREMENT RMS Power (Total or Fundamental) W, VA, VAR Ranges: V∗1 V∗10 V∗100 V∗1000 0.1 m

Page 16

Fluke 433/434 Users Manual 21-10 PF, DPF, COSΦ: Measurement Measurement Range Resolution Accuracy (excl. Clamp error) Power Factor 0.00 ... 1.00

Page 17 - Features Of Fluke 433/434

Specifications Trend Recording21 21-11 Energy Whr, VAhr, VARhr Ranges: V∗1 V∗10 V∗100 V∗1000 0.1 mV/A 000.0 kWhr ... 200.0 GWhr 0.000 MWhr ...

Page 18

Fluke 433/434 Users Manual 21-12 Dips and Swells Item Specification Resolution 25ms, 50ms, 100ms, 200ms, 500ms, 1s, ... , 450d Duration 90s, 180s, 6

Page 19 - Chapter 4

Specifications Display21 21-13 Display Item Specification Additional Information Type LCD color ¼ VGA Liquid Crystal Display color version Vi

Page 20 - Powering the Analyzer

Fluke 433/434 Users Manual 21-14 Printers and Interface Item Specification Additional Information Type RS-232, optically isolated To be used with inte

Page 21 - Locking the keyboard

Specifications Mechanical21 21-15 Mechanical Item Customer Specification Additional Information Height x Width x Depth 256 x 169 x 64 mm 10.

Page 22 - Reset to Factory Defaults

Fluke 433/434 Users Manual 21-16 Electro Magnetic Compatibility (EMC) Item Customer Specification Additional Information Emission and Immunity EN-

Page 23 - Display Information

1 Index —A— A range, 18-4 Accessories, 1-1 Active power, 11-2 Apparent power, 11-2 —B— Banana Inputs, 6-1 Bar Graph screen, 5-2 Battery Charger, 1

Page 24 - Screen Types

Fluke 433/434 Users Manual 1-6 See explanation in manual Double Insulation (Protection Class) Disposal information Earth Recycling information

Page 25

Fluke 433/434 Users Manual 2 Harmonics, 10-1 Hx, 16-5 Hysteresis, 9-1, 15-2 —I— Inductive load, 11-2 Inputs, 6-1 Inrush currents, 15-1 Inrush time,

Page 26

Index (continued) 3 Time, 5-3, 18-3 Transients, 14-1 Trend screen, 5-2 Trigger conditions, 18-7 Troubleshooting, 20-3 —U— U, Unstable, 5-3 Unba

Page 27 - Input Connections

Fluke 433/434 Users Manual 2

Page 28

General Aspects Safety Information: Read First 1 1-7 Slider switch on BC430 Battery Charger / Power Adapter to select line power voltage: 115V

Page 29 - Input Connections 6

Fluke 433/434 Users Manual 1-8

Page 30

2-1 Chapter 2 About This Manual Introduction This Users Manual gives full and comprehensive information on how to use the Fluke 433 and 434 Three P

Page 31 - Scope Waveform and Phasor

Fluke 433/434 Users Manual 2-2 • Chapter 18. Setting up the Analyzer: a comprehensive explanation of adjustments to customize measurements. • Chapte

Page 32 - Tips and Hints

3-1 Chapter 3 Features Of Fluke 433/434 Introduction The Analyzer offers an extensive and powerful set of measurements to check power distribution

Page 33 - Volts/Amps/Hertz

Fluke 433/434 Users Manual 3-2 Measuring modes to investigate details Phase voltages. Should be close to the nominal value. Voltage waveforms must be

Page 34

4-1 Chapter 4 Basic Operations and Menu Navigation Introduction This chapter deals with a number of general aspects of the Analyzer’s operation: •

Page 36

Fluke 433/434 Users Manual 4-2 A hang strap is supplied with the Analyzer. The figure below shows how to attach the strap correctly to the Analyzer.

Page 37 - Dips & Swells

Basic Operations and Menu Navigation Display Brightness 4 4-3 Caution To prevent decrease of battery capacity, charge it at least twice a year.

Page 38

Fluke 433/434 Users Manual 4-4 Menu Navigation Most of the Analyzer functions are menu operated. Arrow keys are used to navigate through menus. The Fu

Page 39

5-1 Chapter 5 Display Information Introduction The Analyzer uses five different screen types to present measuring results in the most effective way

Page 40 - Events Tables

Fluke 433/434 Users Manual 5-2 Phase Colors Measuring results belonging to different phases are presented with individual colors. If for a certain pha

Page 41

Display Information Screen information common for all screen types 5 5-3 C Status indicators. The following symbols may appear on the screen t

Page 42

Fluke 433/434 Users Manual 5-4

Page 43 - Harmonics

6-1 Chapter 6 Input Connections Introduction This chapter explains how to make connection to the power distribution system under test and how to ad

Page 44

Fluke 433/434 Users Manual 6-2 Figure 6-1. Mounting the decals for voltage and current inputs De-energize power systems before making connections whe

Page 45 - 10-3

Input Connections Input Connections 6 6-3 Before making any measurements, set the Analyzer up for the line voltage, frequency, and wiring confi

Page 46

i Table of Contents Chapter Title Page General Aspects...

Page 47 - Power & Energy

Fluke 433/434 Users Manual 6-4

Page 48

7-1 Chapter 7 Scope Waveform and Phasor Introduction Scope mode shows voltages and currents in the power system under test by means of waveforms or

Page 49 - 11-3

Fluke 433/434 Users Manual 7-2 F4 PERSISTence ON memorizes all waveform variations on screen. F5 Switch between HOLD and RUN of screen update. Cursor

Page 50

8-1 Chapter 8 Volts/Amps/Hertz Introduction Volts/Amps/Hertz displays a table with important numerical measuring values. The related Trend screen s

Page 51 - Flicker

Fluke 433/434 Users Manual 8-2 Available function keys: F4 Access to the Trend screen. For description see below. F5 Switch between HOLD and RUN of sc

Page 52

Volts/Amps/Hertz Tips and Hints 8 8-3 Tips and Hints Voltage and frequency should be close to the nominal values of for example 120 V, 230 V, 4

Page 53 - 12-3

Fluke 433/434 Users Manual 8-4

Page 54

9-1 Chapter 9 Dips & Swells Introduction Dips & Swells records Dips, Interruptions, Rapid Voltage Changes, and Swells. Dips (Sags) and Swel

Page 55 - Unbalance

Fluke 433/434 Users Manual 9-2 Figure 9-2. Characteristics of a voltage swell During an Interruption the voltage sinks well below its nominal value.

Page 56

Dips & Swells Trend 9 9-3 VOLTAGETOLERANCEVOLTAGETOLERANCESTEADY TIMEVOLTAGESTEPTRANSITIONTIMETIMEVOLTAGE STEPTRANSITION TIMESTEADY TIMERAP

Page 57

Fluke 433/434 Users Manual ii Screen Types...5

Page 58

Fluke 433/434 Users Manual 9-4 The screen builds up from the right side of the screen and the corresponding values are displayed in the screen header.

Page 59 - Transients

Dips & Swells Tips and Hints 9 9-5 The following Abbreviations and Symbols are used in the tables: Abbreviation Description Symbol Descri

Page 60

Fluke 433/434 Users Manual 9-6

Page 61

10-1 Chapter 10 Harmonics Introduction Harmonics measures and records harmonics and interharmonics (Fluke 434) up to the 50th. Related data such as

Page 62

Fluke 433/434 Users Manual 10-2 e ENTER The Bar Graph display shows the percentage contribution of each of the components related to the full signa

Page 63 - Inrush Currents

Harmonics Table10 10-3 Table To access the Harmonics Table screen: f F3 The table display shows 8 measurements per phase. Using the SETUP ke

Page 64

Fluke 433/434 Users Manual 10-4 F3 Assign arrow keys to Cursor or vertical Zoom operation. F4 Return to table screen. F5 Switch between HOLD and RUN o

Page 65 - 15-3

11-1 Chapter 11 Power & Energy Introduction Power & Energy displays a table with all important power parameters. The related Trend screen s

Page 66

Fluke 433/434 Users Manual 11-2 e ENTER The table displays power data for each phase and in total: real or active power (kW), apparent power (kVA,

Page 67 - Power Quality Monitoring

Power & Energy Trend11 11-3 To access the Energy popup table: f F3 Available function keys: F2 Pulse count mode on/off. F3 Switch Energy

Page 68

Contents (continued) iii Transients ...

Page 69 - 16-3

Fluke 433/434 Users Manual 11-4 PREFerences. With the demand interval set to OFF the Trend functions as usual with automatic horizontal scaling. Avail

Page 70 - Power Quality Main Screen

12-1 Chapter 12 Flicker Introduction Flicker quantifies the luminance fluctuation of lamps caused by supply voltage variations. The algorithm behin

Page 71 - Events Table

Fluke 433/434 Users Manual 12-2 d e ENTER Flicker is characterized by: short term severity Pst (measured over 1 min for fast feedback), short te

Page 72 - Trend Display

Flicker Trend12 12-3 Trend To access Flicker Trend screen: g F4 The parameters in the table update over time. They are recorded whenever the

Page 73 - Bar Graph Screen

Fluke 433/434 Users Manual 12-4 Tips and Hints Use the PF5 flicker trend and half-cycle voltage or current trends to find the source of flicker. Press

Page 74

13-1 Chapter 13 Unbalance Introduction Unbalance displays phase relations between voltages and currents. Measuring results are based upon the funda

Page 75 - Cursor and Zoom

Fluke 433/434 Users Manual 13-2 The table screen shows all relevant numerical values: negative voltage unbalance percentage, zero sequence voltage unb

Page 76 - Cursor on Trend Displays

Unbalance Phasor13 13-3 Cursor. When the Cursor is on, the Trend values at the Cursor are displayed in the screen header. Moving the Cursor off

Page 77 - 17-3

Fluke 433/434 Users Manual 13-4 unbalance should not exceed 10 %. In case of too high unbalance, use other measuring modes to further analyze the powe

Page 78 - Cursor on Bar graph Displays

14-1 Chapter 14 Transients Introduction The Fluke 434 Analyzer can capture waveforms at high-resolution during a variety of disturbances. The Analy

Page 79 - Setting up the Analyzer

Fluke 433/434 Users Manual iv Memory...

Page 80

Fluke 433/434 Users Manual 14-2 e ENTER f F3 g F5 In the Start menu you can choose a trigger event or a combination of trigger events, trans

Page 81 - General Settings

Transients Tips and Hints14 14-3 F3 Assign up/down arrow keys to browse through all captured screens. F4 Persistence on memorizes all waveform v

Page 82

Fluke 433/434 Users Manual 14-4

Page 83 - 18-5

15-1 Chapter 15 Inrush Currents Introduction Inrush Currents can be captured by Fluke 434. Inrush Currents are surge currents that occur when a lar

Page 84

Fluke 433/434 Users Manual 15-2 e ENTER f F3 Use the arrow keys in the Start menu to adjust the trigger limits: expected inrush time, maximum c

Page 85 - FUNCTION PREFerences

Inrush Currents Inrush Trend Display15 15-3 THRESHOLD(= TRIGGER)} HYSTERESISpretriggerDURATIONinrush timet35 s1 12.3 A2 12.3 A3 1.5 A

Page 86

Fluke 433/434 Users Manual 15-4 Tips and Hints Check the peak currents and their duration. Use the Cursor for readout of momentary values. Check if fu

Page 87 - 18-9

16-1 Chapter 16 Power Quality Monitoring Introduction Power Quality Monitoring or System Monitor displays a Bar graph screen. This screen shows whe

Page 88

Fluke 433/434 Users Manual 16-2 12345A / L1B / L2C / L3A / L1B / L2C / L3A / L1B / L2C / L3DIPSINTERRUPTIONSRAPID VOLTAGE CHANGESSWEELSUNBALANCEFREQUE

Page 89 - USER PREFerences

Power Quality Monitoring Introduction16 16-3 average from these measuring values across 10-minute observation periods. The 10-minute averages ar

Page 90

1-1 Chapter 1 General Aspects Introduction This chapter informs you about a number of general and important aspects concerning the Fluke 433/434 Th

Page 91 - Limits Adjustments

Fluke 433/434 Users Manual 16-4 Power Quality Main Screen To access the Power Quality Main screen: c MONITOR d F5 Power Quality Monitoring is re

Page 92

Power Quality Monitoring Events Table16 16-5 Events Table Figure 16-2. Events Table The events table shows the events that occurred during the

Page 93 - 18-15

Fluke 433/434 Users Manual 16-6 Two ways to access Trend: 1. Use the up/down arrow keys to highlight an event in the table. To access Trend press the

Page 94

Power Quality Monitoring Bar Graph Screen16 16-7 Bar Graph Screen Figure 16-4. Bar Graph Screen The main system monitor display shows the worst

Page 95 - Chapter 19

Fluke 433/434 Users Manual 16-8

Page 96

17-1 Chapter 17 Cursor and Zoom Introduction This chapter explains how to use Cursor and Zoom to display and investigate details of Waveform, Tren

Page 97 - Use of Printer and PC

Fluke 433/434 Users Manual 17-2 Figure 17-3. Waveform display with cursor and zoom on Press Function key F2 to obtain a subset with keys to control C

Page 98

Cursor and Zoom From Events Table to Trend Display with Cursor On.17 17-3 Figure 17-6. Trend display with cursor and zoom on The Function keys

Page 99 - Tips and Maintenance

Fluke 433/434 Users Manual 17-4 d F5 Press F5 to obtain trend display with the cursor on and positioned on the highlighted event in the table. Curs

Page 100 - Parts and Accessories

18-1 Chapter 18 Setting up the Analyzer Introduction The SETUP key accesses menus to view and change Analyzer settings. At delivery the Analyzer is

Page 101 - Troubleshooting

Fluke 433/434 Users Manual 1-2 Limited Warranty & Limitation of Liability Each Fluke product is warranted to be free from defects in material and

Page 102

Fluke 433/434 Users Manual 18-2 Figure 18-1. Welcome screen at power-on The Settings are grouped in four functional sections and are explained accord

Page 103 - Specifications

Setting up the Analyzer General Settings18 18-3 General Settings To access the General Settings menus: 1SETUPSETUPENTER1.2345678F1 F2 F3 F4 F510

Page 104 - Electrical Measurements

Fluke 433/434 Users Manual 18-4 6 Freq: adjustment of Nominal Frequency. Use the up/down arrow keys to select 60 or 50 Hz. Press Function key F5 – OK

Page 105 - 21-3

Setting up the Analyzer General Settings18 18-5 Below you will find a step-by-step example on how to change wiring configuration to 3-phase Wye

Page 106

Fluke 433/434 Users Manual 18-6 g F5 An information screen shows up that informs you on how to connect the Analyzer to the power system under test.

Page 107 - 21-5

Setting up the Analyzer FUNCTION PREFerences18 18-7 FUNCTION PREFerences To access the FUNCTION PREFerences menus: SETUPSETUP1.F3ENTER2.3.F1 F3

Page 108

Fluke 433/434 Users Manual 18-8 Measuring Function/ Screen Type Measuring Data to be adjusted Settings Type 1. Scope Waveform Volt, Amp (separate fo

Page 109 - 21-7

Setting up the Analyzer FUNCTION PREFerences18 18-9 The example below shows stepwise how to adjust offset and span of a Volts/Amps/Hertz trend a

Page 110 - Current Unbalance

Fluke 433/434 Users Manual 18-10 h Use the left/right arrow keys to decrease the Voltage offset. i Use the up/down arrow keys to select Volt spa

Page 111 - 21-9

Setting up the Analyzer USER PREFerences18 18-11 USER PREFerences To access the USER PREFerences menus: SETUPSETUP1.ENTER3.F1 F3 F46 7 8 9F4F2

Page 112 - Harmonic Power (Watts only)

General Aspects Declaration of Conformity 1 1-3 Declaration of Conformity Declaration of Conformity for Fluke 433/434 Three Phase Power Qualit

Page 113 - Trend Recording

Fluke 433/434 Users Manual 18-12 2 Phase Colors: use the function keys F1 ... F4 to choose colors as used in the USA, EU, UK, or according to IEC. Or

Page 114 - Wiring Combinations

Setting up the Analyzer Limits Adjustments18 18-13 Limits Adjustments To navigate the Limits Setup menus: SETUPSETUP1.1.F1 F3 F4 F5167F22.0...9A

Page 115 - Display

Fluke 433/434 Users Manual 18-14 Limits Adjustments is used to save, recall, and define sets of limits for: • Power Quality Monitoring. • Dips/Inter

Page 116 - Printers and Interface

Setting up the Analyzer Limits Adjustments18 18-15 6 View Monitor Limits menu. This menu has the same structure as the Edit Monitor Limits menu

Page 117 - Environmental

Fluke 433/434 Users Manual 18-16

Page 118

19-1 Chapter 19 Using Memory, Printer, and PC Introduction This chapter explains how to save screens and data into the Analyzer’s memory and how to

Page 119 - Change wiring config, 18-5

Fluke 433/434 Users Manual 19-2 Memory Operations The MEMORY button accesses menus to save, recall, view, delete and print datasets and screenshots. W

Page 120 - Optional parts, 20-3

Using Memory, Printer, and PC Use of Printer and PC19 19-3 Available function keys for recall and delete: F1 Return to main menu. F2 Access to t

Page 121 - (continued)

Fluke 433/434 Users Manual 19-4 the up/down arrow keys and ENTER. Then adjust the baudrate with the left/right arrow keys and leave the menu with F5 -

Page 122 - Users Manual

20-1 Chapter 20 Tips and Maintenance Introduction This chapter covers basic maintenance procedures that can be performed by the user. For complete

Modèles reliés 433

Commentaires sur ces manuels

Pas de commentaire