Fluke DDS Function Generator 271 Manuel d'utilisateur Page 31

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 34
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 30
Getting Started
Using the Instrument
step name start (µs)
amplitude
start
scaled to 0-1024
step
length
amplitude
scaled to 0-511
0 0 30% 0 29 153
1 sync pulse 1.8 0% 29 75 0
2 back porch 6.5 30% 104 90 153
3 50% grey 12.1 65% 194 165 332
4 white 22.4 100% 358 165 511
5 50% grey 32.8 65% 525 165 332
6 black 43.1 30% 690 165 153
7 50% grey 53.4 65% 854 165 332
8 front porch 63.7 30% 1019 5 153
These values can now be entered in the staircase set-up
screens. Press
EDIT then the STAIR key. You should
see this screen:
VALS=ABS AUTO=YES
STEP=00 ACTIVE
LENGTH=0256
LEVEL=+511
Change the length for the first step (00) to 29 and the level
to +153. If you use the numeric and
ENTER keys to make
the change, the step number automatically increments
after the level has been entered.
21
Vue de la page 30
1 2 ... 26 27 28 29 30 31 32 33 34

Commentaires sur ces manuels

Pas de commentaire