Fluke 87 III Manuel d'utilisateur Page 56

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 58
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 55
80 Series III
Users Manual
48
Table 17. Frequency Counter Sensitivity and Trigger Levels
Minimum Sensitivity (RMS Sinewave) Approximate Trigger Level
Input Range
1
5 Hz - 20 kHz 0.5 Hz - 200 kHz (DC Voltage Function)
400 mV dc
400 mV dc
4 V
40 V
400 V
1000 V
70 mV (to 400 Hz)
150 mV
0.3 V
3 V
30 V
300 V
70 mV (to 400 Hz)
150 mV
0.7 V
7 V (140 kHz)
70 V (14.0 kHz)
700 V (1.4 kHz)
40 mV
1.7 V
4 V
40 V
400 V
Duty Cycle Range Accuracy
0.0 to 99.9% Within ±(0.05% per kHz + 0.1%) of full scale for a 5 V logic family input on the 4 V dc range.
Within ±((0.06 x Voltage Range/Input Voltage) x 100%) of full scale for sine wave inputs on ac voltage ranges.
1. Maximum input for specified accuracy = 10X Range or 1000 V.
Vue de la page 55