Table 19. MIN MAX Recording Specifications
Model Nominal Response Accuracy
83 100 ms to 80%
1 s
Specified accuracy ±12 counts for changes >200 ms in duration (±40 counts in ac with beeper on)
Same as specified accuracy for changes >2 seconds in duration (±40 counts in ac with beeper on)
85, 87 100 ms to 80%
(DC functions)
120 ms to 80%
(AC functions)
1 s
250 µs
(Model 87 only)
Specified accuracy ±12 counts for changes >200 ms in duration
Specified accuracy ±40 counts for changes >350 ms and inputs >25% of range
Same as specified accuracy for changes >2 seconds in duration
Specified accuracy ±100 counts for changes >250 µs in duration
(± 250 digits typical for mV, 400 µA dc, 40 mA dc, 4000 mA dc)