COMMAND REFERENCE 4 - 17
*
ESE Event Status Enable
Syntax:
*
ESE <numeric_data>
Query form:
*
ESE?
Response:
<integer>
Description:
The command sets and the query reports the contents of the standard Event
Status Enable register (ESE). The range of the 8-bit ESE contents is between 0
and 255 decimal. The contents of the standard Event Status Enable (ESE)
register determine which bits in the standard Event Status Register (ESR) are
enabled to be summarized in the Status byte Register (STB). The contents of the
standard ESE register are cleared at Power on.
Example:
Send
*ESE 17 Enables the EXE (Execution Error) and the OPC (Oper-
ation Complete) bits to be summarized in the Status
Byte Register. Alternative commands
*
ESE #B10001
and
*
ESE #H11.
Send
*ESE?
Read
17 The bits 4 (= EXE bit) and 0 (=OPC bit) are set.