Fluke PM-3370B Manuel d'utilisateur Page 56

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 297
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 55
USING THE COMBISCOPE INSTRUMENTS 3 - 37
The following diagram shows the possible states of the acquisition process when
"averaging" is on, and the way they are affected by commands.
Figure 3.13 The Trigger Model during acquisition averaging
INIT
or
INIT:CONT ON
*RST
ABORt
power on
No
Yes
IDLE state
INITiated state
INIT:CONT ON
No
Yes
Wait for completeWait for trigger
1
acquisition
+
averaging
TRIGger
:LEVel
:SLOPe
TRIGger
:SOURce
LINE
INTernal
IMMediate
ST7189
SENSe:AVERage:COUNt
No
Wait for AVERage state
Yes
Vue de la page 55
1 2 ... 51 52 53 54 55 56 57 58 59 60 61 ... 296 297

Commentaires sur ces manuels

Pas de commentaire