Fluke IMPULSE 6000D Manuel d'utilisateur Page 5

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 11
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 4
Manual Supplement Impulse 6000D/7000DP Users
4 3/09
Ventricular Fibrillation ......................................... Coarse and fine
Monomorphic Ventricular Tachycardia ............... 120 to 300 (by 5) BPM
Polymorphic Ventricular Tachycardia ................. 5 types
Asystole .............................................................. Flat line
ECG Waves
ECG General
Lead configuration .............................................. 12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance ..................................... 1000 (nominal)
Rate accuracy..................................................... ±1 % of nominal
ECG Amplitudes
Reference Lead...................................................Selectable, Lead II (default) or Lead I
Settings................................................................0.05 to 0.45 (by 0.05) mV
0.5 to 5.0 (by 0.5) mV
Accuracy (All Performance waves and Normal Sinus R waves)
Lead II..................................................................±2 %
All other leads......................................................±5 %
Defib paddles.......................................................±5 %
Amplitude of ECG signals relative to amplitude setting (in percent)
Lead II reference
Performance waves and R wave detection
Lead # I II III V1 V2 V3 V4 V5 V6
Ref. Amp. 70 % 100 % 30 % 100 % 100 % 100 % 100 % 100 % 100 %
Vue de la page 4
1 2 3 4 5 6 7 8 9 10 11

Commentaires sur ces manuels

Pas de commentaire