Impulse 6000D/7000DP Users Manual Supplement
4/08 7
Accuracy ......................................................... ±(10 % of setting + 0.2 mV)
Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent)
Lead II reference
Lead # I II III V1 V2 V3 V4 V5 V6
Ref. Amp. 67 % 100 % 33 % 67 % 67 % 67 % 67 % 67 % 67 %
Lead I reference
Lead # I II III V1 V2 V3 V4 V5 V6
Ref. Amp. 100 % 150 % 50 % 100 % 100 % 100 % 100 % 100 % 100 %
Arrhythmia Selections
Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)
Demand .......................................................... 30 to 360 (by 1) BPM
Asynchronous
Non-Capture
Non-Function
Threshold (Interactive pacing simulation only) 10 to 250 (by 10) mA
Supraventricular
Atrial Fibrillation Coarse
Atrial Fibrillation fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia