Fluke PM-3370B Manuel d'utilisateur Page 131

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 297
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 130
COMMAND REFERENCE 4 - 31
ABORt
Syntax:
ABORt
Description:
The ABORt command resets the trigger system and places it in the "IDLE" state.
Pending actions that were already started are finished immediately. The ABORt
command is not finished until the pending actions have been terminated.
Note: The commands
*
RST and ABORt have the same effect on the trigger
functions, except that ABORt does not affect the state of the
INITiate:CONTinuous command. So, when an ABORt command is sent
while the INITiate:CONTinuous is ON, the trigger system will leave the
IDLE state at once.
Example:
Send
ABORt Aborts the current acquisition.
Send
CONFigure:AC Configures for AC-RMS value.
Send
READ:AC? Initiates and reads the AC-RMS value.
Read
<the measured AC-RMS value>
Vue de la page 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 296 297

Commentaires sur ces manuels

Pas de commentaire