COMMAND REFERENCE 4 - 117
TRIGger[:SEQuence[1]]:FILTer:LPASs:FREQuency
TRIGger[:STARt]:FILTer:LPASs:FREQuency
TRIGger[:SEQuence[1]]:FILTer:LPASs:STATe
TRIGger[:STARt]:FILTer:LPASs:STATe
Syntax:
TRIGger[:SEQuence[1]]:FILTer:LPASs:FREQuency <NRf>
| MINimum | MAXimum
TRIGger[:SEQuence[1]]:FILTer:LPASs:STATe <Boolean>
Alias:
TRIGger[:STARt]:FILTer:LPASs:FREQuency <NRf>
| MINimum | MAXimum
TRIGger[:STARt]:FILTer:LPASs:STATe <Boolean>
<NRf> The cutoff frequency expressed in hertz. Possible
values are:
- 0 Defines trigger DC coupling (MINimum).
- 10 Defines trigger AC coupling.
- 30000 Defines LF-reject (MAXimum).
<Boolean> 0 | OFF Sets low-pass filter off.
1 | ON Sets low-pass filter on.
Query form:
TRIGger[:SEQuence[1]]:FILTer:LPASs:FREQuency?
MINimum | MAXimum
Alias:
TRIGger[:STARt]:FILTer:LPASs:FREQuency?
MINimum | MAXimum
Response:
0.00E+00 | 1.00E+01 | 3.0E+04
If MINimum was specified, the minimum possible cutoff frequency
is returned, i.e., 0 Hz. If MAXimum was specified, the maximum
possible cutoff frequency is returned, i.e., 30 KHz.
Query form:
TRIGger[:SEQuence[1]]:FILTer:LPASs:STATe?
Alias:
TRIGger[:SEQuence[1]]:FILTer:LPASs:STATe?
Response:
0 | 1
0 High-pass filter active (HF-reject).
1 Low-pass filter active.