701/702
Users Manual
12
Getting Acquainted with Your
Calibrator
Input and Output Jacks
Figure 4 shows the calibrator input and output jacks.
Table 2 explains their use.
Table 2. Input/Output Jacks and Connectors
ITEM FEATURE DESCRIPTION
1 Battery Eliminator jack Jack for the Model BE9005 Battery Eliminator. Use the battery eliminator for bench-
top applications where ac line power is available. This input does not charge the
battery. The battery can only be charged in the external charger.
2 wSERIAL PORT (702
Only)
Connects the calibrator to a personal computer. Instructions for communicating with the
calibrator are in the PMLink Software manual.
3 Pressure module
connector
Connects the calibrator to a pressure module.
4 TC input/output Jack for measuring or simulating thermocouples. This jack accepts a miniature
polarized thermocouple plug with flat, in-line blades spaced 7.9 mm (0.312 in) center to
center.
5, 6 w MEAS V jacks Input jacks for measuring voltage, frequency, or three- or four-wire RTDs (Resistance
Temperature Detectors).
7, 8 w SOURCE mA,
MEAS mA RTD
jacks
Jacks for sourcing or measuring current, measuring resistance and RTDs, and
supplying loop power.
9,10 w SOURCE V RTD
jacks
Output jacks for sourcing voltage, resistance, frequency, and for simulating RTDs
Vue de la page 21