701/702
Users Manual
54
Sourcing in Percent of Scale
To adjust the source value as a percent of scale as
shown below, proceed as follows:
1. If necessary, press M for SOURCE mode.
2. Select a source function (m, v, q, t,
or p) as previously described.
3. Press the Scale softkey.
4. Use the numeric keypad to enter the 0% of scale
value.
5. Press e.
6. Use the numeric keypad to enter the 100% of
scale value.
7. Press e.
8. Press the Done softkey.
Percent of scale remains in effect until you press the
Scale softkey again.
Stepping and Ramping the Output Value
Two special features are available for adjusting the
value of source functions except pressure (sourcing
pressure requires that you use an external pressure
source):
Stepping the output with the
u and d keys.
Ramping the output with optional continuity or V
trip detect.
Stepping the Output
The step feature allows you to select a step size in
engineering units (mV, V, mA, °C, etc.) or % of scale.
Stepping the output in % of scale is useful for quickly
jumping between 0% and 100% (set step size =
100%) or 0-50-100% (set step size = 50%). Stepping