701/702
Users Manual
50
Simulating Thermocouples
NOTE
Refer to “Measuring Temperature” earlier in the
manual for a table of data relating to
thermocouple types supported by the calibrator.
Connect the calibrator TC connector to the
instrument under test with thermocouple wireand
theappropriate thermocouple mini-connector
(polarized thermocouple plug with flat, in-line blades
spaced 7.9 mm (0.312 in) center to center).
One pin
is wider than the other. Do not try to force a miniplug
in the wrong polarization
. Figure 18 shows this
connection. Proceed as follows to simulate a
thermocouple:
1. Attach the thermocouple leads to the appropriate
TC miniplug, then to the TC jack as Figure 12
shows.
2. If necessary, press M for SOURCE mode.
3. Press tfor the display that prompts you to
enter thermocouple type.
4. Press the u or d key followed by e to
select the desired thermocouple type.
5. Enter the temperature you want to simulate as
prompted by the display and press e.
NOTE
If you use copper wire instead of thermocouple wire,
the reference junction is no longer inside the
calibrator. The reference junction is moved to the
instrument (transmitter, indicator, controller, etc.)
input terminals. You must measure this external
reference temperature accurately and enter it into the
calibrator. Do this by pressing s and setting
Ref.
Junc. Compensat
and Ref. Junc. Temp. After you
enter the external reference temperature, the
calibrator corrects all voltages to compensate for this
new reference junction temperature.