Vue de la page 82
Documenting Process Calibrator
Quick Guide to Applications
73
the list of tasks (procedures) downloaded from
PMLink. Tasks are 702 calibrator configurations,
saved with a procedure name, for example the type
and manufacturer of a specific transmitter. A
procedure configures the calibrator for transmitter
calibration with all the calibration parameters
(SOURCE/MEASURE functions, 0% and 100%
levels, test strategy) predefined. See the
PMLink
manual for instructions on developing and
downloading tasks.
Clearing Memory
The Clear Memory softkey, shown after you press
More Choices softkey three times, clears all the
memory:
Saved results
Min Max data
Log data sets (Model 702 only)
A confirmation message appears so that you do not
inadvertently erase the memory.
Quick Guide to Applications
The following figures show test lead connections and
which calibrator function to use for many different
applications.