Vue de la page 77
701/702
Users Manual
68
Memory Operations
Saving Results
As Found/As Left test results are automatically saved
at the end of each test routine. Any other time during
MEASURE, SOURCE, or MEASURE/SOURCE you
can press the Save softkey to save the data on the
display for later review.
After you press Save, the calibrator saves the
information on the display and shows a saved result
index number, the date and time, and the cumulative
percentage of memory used, as in the following
display:
You can quit at this point. Or, if you want to add
information to the saved data, the calibrator has a
way for you to do so. If you press the Continue
softkey, the display prompts you to enter the
instrument tag identifier (Tag), instrument serial
number (Ser#), and operator name (ID), as shown in
the following display: